Acoustic Tweezers and Acoustofluidic Cell Sorting Statistics (2026): 55+ Data Points on Sorting Throughput, Cell Viability, and CTC and Exosome Isolation

Acoustic tweezers and acoustofluidic cell sorting statistics 2026: 63 CTC-containing clusters found in 9 of 12 prostate cancer patients, versus none by CellSearch.

Acoustic enrichment found 63 CTC-containing cell clusters in 9 of 12 men with metastatic prostate cancer, while the CellSearch comparator system detected no clusters at all in the same patients’ blood (Magnusson et al., Analytical Chemistry 2024). The same sound-based toolbox removed more than 99.999% of blood cells before isolating exosomes at 98.4% purity (Wu et al., PNAS 2017), and a 2024 acoustic trapping method reported 93% extracellular vesicle yield in 1 minute against 50% in 480 minutes for ultracentrifugation (Rufo et al., Microsystems and Nanoengineering 2024). Speed is still the open question: the fastest acoustic sorters reported here run in the low thousands of events per second, below commercial FACS (see section 7). We aggregated data from Nature Methods, PNAS, Analytical Chemistry, Lab on a Chip, Science Advances, Scientific Reports, Cytotherapy, Clinical Chemistry and Laboratory Medicine, and review articles indexed in Europe PMC, with a narrower focus on cell sorting in flowing microchannels than our roundup of acoustic levitation and ultrasonic particle manipulation.

A note on vocabulary used throughout: “acoustic tweezers” are devices that use sound waves to push, trap or sort cells and particles without touching them. “Purity” is the share of the collected output that is the target cell type; “recovery” (or yield) is the share of the target cells in the input that end up in the output. A good device needs both.

TL;DR

  • Absorbance-activated acoustic droplet sorting reached 1 kHz (1,000 droplets per second) with 98.5% recovery and 99.9% purity (Richter et al., Lab on a Chip 2022).
  • A tilted-angle surface acoustic wave chip processed circulating tumor cell samples at 7.5 mL/h, up from about 1.2 mL/h in earlier work (Wu et al., Small 2018).
  • Acoustic washing recovered 99% of red blood cells while removing 99% or more of albumin with a two-step loop wash (Sugier et al., Cytotherapy 2023).
  • Acoustophoretic plasma separation left less than 0.01% of cells in the plasma, with analyte differences versus centrifuged plasma of 0.84% to 10.50% (Nath et al., Clinical Chemistry and Laboratory Medicine 2025).
  • HeLa cell viability was 90.4% with the acoustic field on versus 91.5% with it off (Li et al., PNAS 2015).
  • Acoustic enrichment found 63 CTC-containing clusters in 9 of 12 prostate cancer patients; CellSearch found none (Magnusson et al., Analytical Chemistry 2024).
  • An integrated acoustic chip removed more than 99.999% of blood cells and isolated exosomes at 98.4% purity and 82.4% recovery (Wu et al., PNAS 2017).
  • Oscillating microbubble arrays isolated exosomes at 93% purity from undiluted mouse whole blood in about 3 minutes, versus about 16 hours for conventional methods (Li et al., Science Advances 2025).
  • Acoustic tweezers use 0.01 to 10 W/cm2, versus 1,000,000 to 10,000,000 W/cm2 for optical tweezers (Ozcelik et al., Nature Methods 2018).
  • An image-based acoustic sorter reached up to 400 cells per second at more than 92% purity and more than 90% viability (Nawaz et al., Lab on a Chip 2023).
  • FACS sorting raised the oxidized-to-reduced glutathione ratio nearly 7-fold in sorted cells (Llufrio et al., Redox Biology 2018).

1. Throughput: Bulk Waves, Surface Waves and the Flow-Rate Gap

The acoustofluidics field splits into two device families, and the split shows up first in throughput. Bulk acoustic wave (BAW) devices vibrate a whole channel to set up a standing wave across it; surface acoustic wave (SAW) devices send waves along the surface of a piezoelectric chip, which gives finer control at smaller volumes. BAW devices reach about 10 mL/min, while SAW devices handle less than 1 mL/min (Ozcelik et al., Nature Methods 2018). Most of the precision results in this roundup come from SAW chips; most of the volume results come from BAW.

The practical benchmark is a blood tube. A tilted-angle standing SAW chip ran 7.5 mL of patient blood at 7.5 mL/h, which works out to roughly one hour per standard tube (7.5 mL / 7.5 mL/h = 1 h), and the authors describe it as up to a sevenfold improvement over the roughly 1.2 mL/h of the earlier design (Wu et al., Small 2018). Clinical blood processing sets a much higher bar: apheresis requires 30 to 80 mL/min (Hossein and Angeli, Biophysical Reviews 2023), or 1,800 to 4,800 mL/h, which is 240 to 640 times the 7.5 mL/h chip.

MetricValueSource
Typical throughput, bulk acoustic wave (BAW) devices vs surface acoustic wave (SAW) devicesAbout 10 mL/min vs less than 1 mL/minOzcelik et al., Nature Methods 2018
Flow rate range of the AcouWash 2 acoustic cell washer (flow unstable at 900 µL/min)100 to 900 µL/minAdler et al., Scientific Reports 2025
Tumor cell separation throughput, tilted-angle SAW chip, 2018 design vs earlier design7.5 mL/h vs about 1.2 mL/hWu et al., Small 2018
Acoustic cell washing, input flow rate45 mL/hSugier et al., Cytotherapy 2023
Neuroblastoma cell separation from stem cell products, cell throughput100,000 cells/minOlm et al., Scientific Reports 2019
Required flow rate for clinical apheresis (blood component collection)30 to 80 mL/minHossein and Angeli, Biophysical Reviews 2023 (citing Zhou and Papautsky 2020)

Context note: several throughput benchmarks in this section predate 2023; the 2018 tilted-angle SAW figure is the most recent available data for that device class. The 2025 Analytical Chemistry review of the field (Godary et al.) cites one acoustofluidic device operating at 3 to 520 µL/min (Mendis et al.), still well below what apheresis requires.

2. Separation Purity and Recovery: Blood Components and Cell Washing

Acoustic separation works because cells differ in size, density and compressibility, and the acoustic radiation force pushes them toward or away from pressure nodes accordingly. The long-run record on blood components is strong: acoustic separation of platelets from red blood cells reached 99% platelet recovery (Petersson 2007, cited in Wu et al., Microsystems and Nanoengineering 2019). The newer numbers are about whether that precision survives real manufacturing and clinical workflows.

Two recent studies suggest it does, with trade-offs. A cell-washing system for cell therapy production removed up to 90% of albumin in a single pass while recovering 99% of red blood cells, and a two-step loop pushed albumin removal to 99% or more with 99% recovery; washed adipose-derived stromal cells showed only two differentially expressed genes (Sugier et al., Cytotherapy 2023). An in-line plasma separator aimed at reducing blood loss in premature infants left fewer than 0.01% of cells in the plasma, yet still more residual cells than centrifugation (median 642 vs 205 per µL) (Nath et al., Clinical Chemistry and Laboratory Medicine 2025).

MetricValueSource
Separation efficiency for peripheral blood mononuclear cells (lymphocytes and monocytes), at up to 20 µL/min85%Godary et al., Analytical Chemistry 2025 review (citing Alsved et al.)
Lymphocyte and monocyte enrichment from whole blood (fold increase) and recovery2,800-fold; 88%Wu et al., Microsystems and Nanoengineering 2019 review (citing Urbansky et al.)
Albumin removal with red blood cell recovery, single acoustic washing passUp to 90% removal; 99% recoverySugier et al., Cytotherapy 2023
Albumin removal with cell recovery, two-step loop washing99% or more removal; 99% recoverySugier et al., Cytotherapy 2023
Differentially expressed genes in stromal cells after loop washing vs input2 genes (HES4 and MIR-3648-1)Sugier et al., Cytotherapy 2023
Cells remaining in plasma after acoustophoretic separationLess than 0.01%Nath et al., Clinical Chemistry and Laboratory Medicine 2025
Median platelets per µL in plasma, acoustophoresis vs centrifugation20,477 vs 1,537Nath et al., Clinical Chemistry and Laboratory Medicine 2025
Difference in clinical chemistry results, acoustophoresis plasma vs centrifuged plasma (12 tests)0.84% (sodium, 95% CI 0.48-1.19) to 10.50% (AST, 95% CI 5.02-15.99)Nath et al., Clinical Chemistry and Laboratory Medicine 2025

Outlier note: the Nath study also found lower cell-free hemoglobin in acoustically separated plasma (range 0.0-0.2 g/L) than in centrifuged plasma (0.1-0.3 g/L), consistent with less red cell rupture (hemolysis), while platelet carryover was the clear weak point.

3. Cell Viability After Acoustic Exposure

Viability is where acoustic methods make their strongest case. Across sorter designs, frequencies and cell types, post-sort viability in the papers reviewed here sits between about 90% and 99%, and controlled comparisons show no meaningful loss versus unexposed cells. In a tumor cell separation study, viability was 90.4 ± 4.7% with the acoustic field on and 91.5 ± 7.6% with it off (Li et al., PNAS 2015).

The more careful tests look beyond survival. A 2013 study drove microglia through a 1.94 MHz field at up to 20 Vpp (volts peak-to-peak, a measure of drive voltage) and found no significant change in cell death, apoptosis, mitochondrial function or inflammatory response; cells spent about 1 second in the acoustic field (Burguillos et al., PLoS One 2013). Red blood cells held in an acoustic device for up to 30 minutes showed no change in viability (Ozcelik et al., Nature Methods 2018). Most recent available data for the foundational biocompatibility work: Burguillos et al., 2013.

MetricValueSource
HeLa cell viability, acoustic field on vs off90.4 ± 4.7% vs 91.5 ± 7.6%Li et al., PNAS 2015
Microglia cell death, unprocessed control vs 20 Vpp acoustic processing9.7 ± 5.7% vs 11.6 ± 6.9% (no significant difference)Burguillos et al., PLoS One 2013
Post-sort viability, acoustofluidic fluorescence-activated sorter (HeLa cells)99.18%Nawaz et al., Analytical Chemistry 2015
Normalized viability, standing-wave acoustic fluorescence sorter, relative to unsorted cells94.9 ± 2.0%Ren et al., Small 2018
Post-sort viability, acoustic droplet sorter with focused transducers93.5%Zhong et al., Small 2021
Dead cells after two-step acoustic tumor cell separation vs ice-incubated controls3.2% vs 8.2%Undvall Anand et al., Analytical Chemistry 2021
Necrotic cells after acoustic washing vs centrifuge washing (radiolabeled murine T lymphoblast cells)7% ± 2% vs 12% ± 3% (P = 0.0093)Adler et al., Scientific Reports 2025
T-cell proliferation at day 6, acoustically sorted vs non-sorted89.8 ± 5.0% vs 86.7 ± 4.3% (no significant difference)Olm et al., Scientific Reports 2019

Context note: gentleness is not free. The same 2025 study found average cell recovery of 22% ± 10% with the acoustic washer versus 55% ± 2% with the centrifuge (P = 0.06, n = 5), so the cell-friendly method lost more cells (Adler et al., Scientific Reports 2025).

4. Circulating Tumor Cell (CTC) Isolation from Blood

Circulating tumor cells are cancer cells shed into the bloodstream, and they are rare: usually in the range of 1 to 100 cells per mL of blood, hidden among millions of white blood cells (Li et al., PNAS 2015). Acoustic separation is label-free, meaning it does not depend on an antibody binding a surface marker, so it can in principle catch tumor cells that marker-based tests miss. A 2024 patient study is the strongest evidence of that in this roundup: acoustic enrichment found 63 CTC-containing cell clusters across 9 of 12 metastatic prostate cancer patients, and detected more CTCs than CellSearch in all 10 patients where both were run (Magnusson et al., Analytical Chemistry 2024).

The limits are also visible. Patient CTCs ranged from 7 to 28 µm in diameter, so the smallest tumor cells overlap with white blood cells and can be lost, and the cohort was only 12 patients. Purity remains the hard part: a two-step BAW method achieved 459 ± 188-fold white blood cell depletion but recovered 42% of viable cancer cells (Undvall Anand et al., Analytical Chemistry 2021). Information on liquid biopsy methods here is technical, not clinical guidance.

MetricValueSource
Spiked cancer cell recovery, tilted-angle SAW (white blood cell removal about 90%)More than 83% (83% to 96%)Li et al., PNAS 2015
CTCs found in metastatic breast cancer patient blood, first two patients (2 mL each)59 and 8Li et al., PNAS 2015
Spiked cancer cell recovery at 7.5 mL/hMore than 86%Wu et al., Small 2018
CTC counts in 5 castration-resistant prostate cancer patients0.93 to 400 per mLWu et al., Small 2018
Recovery and purity, acoustophoresis of spiked prostate cancer cells at 70 µL/min72.5-93.9% recovery; 79.6-99.7% purityWu et al., Microsystems and Nanoengineering 2019 review (citing Augustsson 2012)
Patients with CTC counts above the healthy-control cutoff (13.8 cells per 6 mL)7 of 12Magnusson et al., Analytical Chemistry 2024
CTC-containing clusters detected, acoustic enrichment vs CellSearch63 clusters in 9 of 12 patients vs noneMagnusson et al., Analytical Chemistry 2024
White blood cell depletion and viable cancer cell recovery, two-step BAW (10,000 spiked cells)459 ± 188-fold; 42%Undvall Anand et al., Analytical Chemistry 2021

Outlier note: median CTC diameter in the 2018 prostate cohort was 16.5 µm versus 11.8 ± 0.54 µm for white blood cells (Wu et al., Small 2018), and the 2024 patient study above reported a median of 17.0 µm. Size-based acoustic sorting depends on that gap.

5. Exosome and Extracellular Vesicle Separation

Exosomes are nanoscale vesicles (roughly 30 to 150 nm) released by cells and carrying proteins and RNA; they are hard to isolate because they are thousands of times smaller than blood cells. The conventional reference, ultracentrifugation, is slow and lossy: Table 1 of a 2023 review lists 5% to 50% yield and 23% to 70% purity for it, versus about 82% yield and 98% purity for acoustofluidics (Hossein and Angeli, Biophysical Reviews 2023). The integrated acoustic chip behind that figure removed more than 99.999% of blood cells and processed 100 µL of whole blood in about 25 minutes (Wu et al., PNAS 2017).

The 2024-2025 work shifts from purity to speed and sample size. One acoustic disc method reported a 30-minute sample-to-answer time from 10 µL of blood, with 86% yield and a 120-fold purity increase (Naquin et al., Science Advances 2024). A 2025 design that its authors call a metamaterial (the same engineered-structure idea covered in our acoustic metamaterials roundup), built from 46,750 oscillating microbubbles, isolated 93% pure exosomes from undiluted mouse blood in about 3 minutes on a chip costing less than $1 (Li et al., Science Advances 2025).

MetricValueSource
Blood cell removal rate, integrated acoustofluidic exosome chipMore than 99.999%Wu et al., PNAS 2017
Exosome purity and recovery from whole blood (particles under 140 nm)98.4% purity; 82.4% recoveryWu et al., PNAS 2017
Exosome yield and purity, ultracentrifugation (literature range)5-50% yield; 23-70% purityHossein and Angeli, Biophysical Reviews 2023
Exosome yield and purity, ExoQuick precipitation kit (literature range)40-80% yield; 28-87% purityHossein and Angeli, Biophysical Reviews 2023
Extracellular vesicle yield and processing time, FLOAT acoustic trapping vs ultracentrifugation (urine)93% vs 50%; 1 min vs 480 minRufo et al., Microsystems and Nanoengineering 2024
Exosome yield and purity increase, ASCENDx acoustic disc86% yield; 120-fold purity increaseNaquin et al., Science Advances 2024
Exosome purity and processing time, oscillating microbubble arrays vs conventional methods93% in about 3 min vs about 16 hoursLi et al., Science Advances 2025
Vesicle yield, standing SAW nanofilter for vesicles under 200 nmAbout 90%Lan et al., Advanced Science 2025 review (citing Lee et al. 2015)

Context note: the diagnostic results attached to these methods come from small sets. ASCENDx reported 95.8% sensitivity and 100% specificity for classifying cancer versus healthy samples, but on spectra from a small patient set (Naquin et al., Science Advances 2024), and the microbubble study used mouse blood only.

6. Operating Frequencies, Power and Forces

Acoustic tweezers occupy a gentle corner of the physics. They apply 0.01 to 10 W/cm2 of acoustic power, versus 1,000,000 to 10,000,000 W/cm2 for optical tweezers, at frequencies from 1 kHz to 500 MHz (Ozcelik et al., Nature Methods 2018). The authors compare this with diagnostic ultrasound at 2 to 18 MHz and under 1 W/cm2; therapeutic ultrasound is a different regime entirely, as our HIFU tumor ablation statistics show. The forces involved are tiny: acoustic radiation forces on red blood cells are approximately 2 pN.

Frequency sets the scale of control. SAW devices used in cell biology generally operate at 10 to 400 MHz, with standing-wave wavelengths of 10 to 400 µm matched to cell sizes (He et al., Science Advances 2025). Power figures below are often given in dBm, a logarithmic unit where 24 dBm is about 251 mW and 33 dBm about 2 W. Transducer design matters as much as frequency: a focused-transducer droplet sorter runs below 20 Vpp, where other sorters are described as needing more than 200 Vpp (Zhong et al., Small 2021). For the hardware market behind these components, see our ultrasound transducer market statistics.

MetricValueSource
Typical operating frequency range, BAW vs SAW devices100 kHz to 10 MHz vs 1 MHz to 1 GHzWu et al., Microsystems and Nanoengineering 2019 review
Input power, acoustic tweezers vs optical tweezers0.01 to 10 W/cm2 vs 1,000,000 to 10,000,000 W/cm2Ozcelik et al., Nature Methods 2018
Drive voltage, focused-transducer acoustic droplet sorter vs other sortersBelow 20 Vpp vs more than 200 VppZhong et al., Small 2021
SAW frequency range used in cell biology and matching wavelength10 to 400 MHz; 10 to 400 µmHe et al., Science Advances 2025
BAW prostate CTC enrichment, prefocusing and separation fieldsAbout 5 MHz and about 2 MHzMagnusson et al., Analytical Chemistry 2024
Tilted-angle SAW CTC separation, frequency and power19.573 MHz; 37-38 dBmLi et al., PNAS 2015
Image-based focused traveling SAW sorter, frequency and power vs earlier standing SAW design135 MHz; about 1.9 W vs about 3.8 WNawaz et al., Lab on a Chip 2023
Absorbance-activated droplet sorter, SAW frequency and power162 MHz; 251 mW (24 dBm)Richter et al., Lab on a Chip 2022

Context note: heat is the usual constraint. The Burguillos study reported that 25 Vpp exceeded its 37 C temperature limit, and the 2025 microbubble array reported a temperature rise of only 1.92 C (Li et al., Science Advances 2025).

7. Acoustic Sorters vs FACS and Other Methods

Fluorescence-activated cell sorting (FACS) is the reference instrument for sorting cells by fluorescent labels, and on raw speed it still wins. Commercial FACS exceeds 10,000 particles per second, while the first all-acoustic FACS system sorted 150 particles per second (Wu et al., Microsystems and Nanoengineering 2019 review). SAW designs have since closed much of the gap: a standing-wave acoustic sorter held purity above 90% at 2,500 events per second, which its authors still describe as around one order of magnitude slower than commercial FACS (Ren et al., Small 2018).

The acoustic argument is about what happens to the cells, not speed. In astrocytes sorted with FACS (45 psi sheath pressure), the oxidized-to-reduced glutathione ratio rose nearly 7-fold, reactive oxygen species rose by about 50%, and the ADP/ATP ratio more than 4-fold (Llufrio et al., Redox Biology 2018). That study did not test acoustic sorting, so it is a FACS stress benchmark rather than a head-to-head result. Newer acoustic designs add capabilities FACS lacks, such as image-based decisions (Nawaz et al., Lab on a Chip 2023) and kHz droplet sorting (Richter et al., Lab on a Chip 2022).

MetricValueSource
Throughput, commercial FACS vs first all-acoustic (BAW) FACS systemMore than 10,000 vs 150 particles/sWu et al., Microsystems and Nanoengineering 2019 review
Throughput and purity, acoustofluidic fluorescence-activated sorter (HeLa cells)About 1,200 events/s; 92.3 ± 3.39%Nawaz et al., Analytical Chemistry 2015
Throughput at which purity stays above 90%, standing-wave acoustic fluorescence sorter2,500 events/sRen et al., Small 2018
Theoretical maximum sorting rate from 100 µs acoustic deflection time10,000 events/sRen et al., Small 2018
Image-based acoustic sorting rate and purityUp to 400 cells/s; more than 92%Nawaz et al., Lab on a Chip 2023
Acoustic droplet sorter throughput and purity (cell-laden droplets, 14.8% before sorting)More than 1,000 events/s; up to 99.2%Zhong et al., Small 2021
Absorbance-activated acoustic droplet sorting rate, recovery and purity1 kHz; 98.5% recovery; 99.9% purityRichter et al., Lab on a Chip 2022
Oxidized-to-reduced glutathione ratio in FACS-sorted vs rapidly processed cellsNearly 7-fold higherLlufrio et al., Redox Biology 2018

Context note: speed comparisons mix label-based and label-free sorting, beads and cells. The 2018 standing-wave sorter above is the most recent available data for a direct acoustic-versus-FACS throughput statement. The 2025 Analytical Chemistry review (Godary et al.) reports a droplet sorter at 1,000 droplets/s and sorting of stem cell-derived beta cell clusters at up to 0.2 clusters/s, showing how far throughput varies with the object being sorted.

Summary: Acoustic Tweezers and Acoustofluidic Cell Sorting by the Numbers

MetricValueSource
CTC-containing clusters, acoustic enrichment vs CellSearch (12 prostate cancer patients)63 in 9 of 12 patients vs noneMagnusson et al., Analytical Chemistry 2024
White blood cell depletion and viable cancer cell recovery, two-step BAW separation459 ± 188-fold; 42%Undvall Anand et al., Analytical Chemistry 2021
Absorbance-activated acoustic droplet sorting rate and purity1 kHz; 99.9%Richter et al., Lab on a Chip 2022
Tumor cell separation throughput, tilted-angle SAW chip7.5 mL/hWu et al., Small 2018
Required flow rate for clinical apheresis30 to 80 mL/minHossein and Angeli, Biophysical Reviews 2023
Albumin removal with cell recovery, two-step acoustic washing99% or more; 99%Sugier et al., Cytotherapy 2023
Cells remaining in plasma after acoustophoretic separationLess than 0.01%Nath et al., Clinical Chemistry and Laboratory Medicine 2025
HeLa cell viability, acoustic field on vs off90.4% vs 91.5%Li et al., PNAS 2015
Post-sort viability, acoustofluidic fluorescence sorter99.18%Nawaz et al., Analytical Chemistry 2015
Necrotic cells, acoustic vs centrifuge washing7% vs 12%Adler et al., Scientific Reports 2025
Cell recovery, acoustic vs centrifuge washing22% vs 55%Adler et al., Scientific Reports 2025
Blood cell removal before exosome isolationMore than 99.999%Wu et al., PNAS 2017
Exosome purity and recovery, acoustofluidic chip98.4%; 82.4%Wu et al., PNAS 2017
Exosome yield, ultracentrifugation (literature range)5% to 50%Hossein and Angeli, Biophysical Reviews 2023
Extracellular vesicle yield and time, FLOAT vs ultracentrifugation93% in 1 min vs 50% in 480 minRufo et al., Microsystems and Nanoengineering 2024
Exosome purity and time, oscillating microbubble arrays93% in about 3 minLi et al., Science Advances 2025
Acoustic vs optical tweezers input power0.01-10 vs 1,000,000-10,000,000 W/cm2Ozcelik et al., Nature Methods 2018
Acoustic sorter throughput with purity above 90%2,500 events/sRen et al., Small 2018
Image-based acoustic sorting rateUp to 400 cells/sNawaz et al., Lab on a Chip 2023
Oxidized-to-reduced glutathione ratio, FACS-sorted vs rapidly processed cellsNearly 7-fold higherLlufrio et al., Redox Biology 2018

Methodology and Sources

We searched Europe PMC and the open web for peer-reviewed acoustofluidic separation studies and reviews, prioritized open-access full texts we could read in full, and, for figures taken from reviews, named the original study the review cites. Every value above was read in the cited paper or review during research for this update; two sources (Sugier 2023, Nath 2025) were read at abstract level only.

  • Magnusson et al., Acoustic Enrichment of Heterogeneous Circulating Tumor Cells and Clusters from Metastatic Prostate Cancer Patients, Analytical Chemistry 2024
  • Ozcelik et al., Acoustic tweezers for the life sciences, Nature Methods 2018
  • Wu et al., Acoustofluidic separation of cells and particles, Microsystems and Nanoengineering 2019
  • Hossein and Angeli, A review of acoustofluidic separation of bioparticles, Biophysical Reviews 2023
  • Godary et al., Acoustofluidics: Technology Advances and Applications from 2022 to 2024, Analytical Chemistry 2025
  • Lan et al., Microfluidic Nanoparticle Separation for Precision Medicine, Advanced Science 2025
  • He et al., Acoustic technologies for the orchestration of cellular functions for therapeutic applications, Science Advances 2025
  • Li et al., Acoustic separation of circulating tumor cells, PNAS 2015
  • Wu et al., Circulating Tumor Cell Phenotyping via High-Throughput Acoustic Separation, Small 2018
  • Undvall Anand et al., Two-Step Acoustophoresis Separation of Live Tumor Cells from Whole Blood, Analytical Chemistry 2021
  • Wu et al., Isolation of exosomes from whole blood by integrating acoustics and microfluidics, PNAS 2017
  • Naquin et al., Acoustic separation and concentration of exosomes for nucleotide detection: ASCENDx, Science Advances 2024
  • Rufo et al., High-yield and rapid isolation of extracellular vesicles by flocculation via orbital acoustic trapping: FLOAT, Microsystems and Nanoengineering 2024
  • Li et al., Oscillating microbubble array-based metamaterials for rapid isolation of high-purity exosomes, Science Advances 2025
  • Burguillos et al., Microchannel acoustophoresis does not impact survival or function of microglia, leukocytes or tumor cells, PLoS One 2013
  • Adler et al., Centrifuge-free cell radiolabeling using acoustophoresis, Scientific Reports 2025
  • Olm et al., Label-free neuroblastoma cell separation from hematopoietic progenitor cell products using acoustophoresis, Scientific Reports 2019
  • Sugier et al., Feasibility of an acoustophoresis-based system for a high-throughput cell washing, Cytotherapy 2023 (abstract)
  • Nath et al., Acoustophoresis-based blood sampling and plasma separation for potentially minimizing sampling-related blood loss, Clinical Chemistry and Laboratory Medicine 2025 (abstract)
  • Nawaz et al., Acoustofluidic Fluorescence Activated Cell Sorter, Analytical Chemistry 2015
  • Ren et al., Standing Surface Acoustic Wave (SSAW)-Based Fluorescence-Activated Cell Sorter, Small 2018
  • Nawaz et al., Image-based cell sorting using focused travelling surface acoustic waves, Lab on a Chip 2023
  • Richter et al., Acoustic sorting of microfluidic droplets at kHz rates using optical absorbance, Lab on a Chip 2022
  • Zhong et al., Acoustofluidic Droplet Sorter Based on Single Phase Focused Transducers, Small 2021
  • Llufrio et al., Sorting cells alters their redox state and cellular metabolome, Redox Biology 2018
  • Data watch: this is a research field, not a market, so there is no industry-wide adoption or market-size figure here, and we did not include one. Many landmark benchmarks (Li 2015, Wu 2017, Ozcelik 2018, Ren 2018) are older than three years and remain the most recent available data for their specific metric. Figures quoted from the 2019 and 2023 reviews come from studies those reviews cite; where a review did not give the original year, we name the original author only. Device results are mostly from spiked cell lines and small patient sets (12 patients in Magnusson 2024, 5 in Wu 2018), so they are not clinical performance figures. Reported purity and recovery depend heavily on spike ratio, flow rate and power, which makes cross-study comparison approximate. A 2026 Analyst paper on a focused traveling SAW chip for red and white blood cell sorting (Deng et al., volume 151) was not accessible in full text and was excluded. The FLOAT paper gives 20 mL in its abstract but 100 mL in its introduction as the ultracentrifugation baseline volume, and the 1,000-cell result in Undvall Anand 2021 reports 247-fold depletion in the abstract but 282-fold in the results; we used neither disputed figure as a headline.
  • Last updated: October 25, 2026. We update this roundup quarterly, and the next refresh is expected when the next Analytical Chemistry acoustofluidics review and new 2026 Lab on a Chip CTC and exosome studies are indexed in Europe PMC.